127
6.1.2 QC Lockout
If the required QC Test is not performed
within the time or number of measurements
indicated by the QC Reminder, FIGURE 6-1,
the Analyzer will perform a QC lockout.
FIGURE 6-3 will be displayed.
FIGURE 6-3
The QC Lockout(1) indicates that the
Analyzer has performed a QC lockout.
Patient Tests are not allowed when the
Analyzer has performed a QC lockout.
Linearity and Proficiency Tests are not
allowed when the Analyzer has performed a
QC lockout.
STAT Tests are not allowed when the
Analyzer has performed a QC lockout if
usage of STAT Tests is set to “Not Used” (see
3.4.2 Patient Test Settings) or if the number
of STAT Tests performed has reached the
max. limit allowed.
The Information text(2) indicates the types
of QC Tests required to unlock the QC
Lockout.
a) To enter the QC Test procedure, press the
QC Test button(3).
FIGURE 6-6 will be displayed.
1 QC Lockout
2 Information text
3 QC Test button
08/09/03 10:55 AM
JAMES SMITH
Run QC 1, QC 2
1
3
2

Table of Contents

Other manuals for HemoCue 201 DM

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HemoCue 201 DM and is the answer not in the manual?

HemoCue 201 DM Specifications

General IconGeneral
BrandHemoCue
Model201 DM
CategoryDocking Station
LanguageEnglish

Related product manuals