EasyManuals Logo
Home>Honeywell>Desktop>enCore FC Series

Honeywell enCore FC Series User Manual

Honeywell enCore FC Series
133 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #102 background imageLoading...
Page #102 background image
Maintenance 8
Page 102 of 120
Operating Instructions
enCore FC
The enCore FC automatically signals as soon as battery state of charge has
dropped to 20
% of that of a new battery.
A dead battery may sometimes lead to a loss of the measurement
data that was calculated and saved.
For safety reasons, the battery should be replaced only by the
manufacturer's service department or by an appropriately trained
specialist working for the plant operator.
The housing of the device must be opened in order to replace
the battery. The presence of a metrology expert may be required
for this purpose when the device is being used within the scope
of legal metrology.
ATTENTION!
Risk of short circuit
The device may damage if it is opened in case of connected
power supply.
Always make sure the enCore FC is free of voltage before
changing the battery.
8.1.1 Changing Batteries
You need one of the below mentioned batteries: Lithium 3 V CR2 ½ AA
For reasons of security, use enSuite to read out the parameterization of
the enCore FC.
Switch the device to free of voltage.
Remove all cable connections to the enCore FC.
Remove the enCore FC from the switch cabinet.

Table of Contents

Other manuals for Honeywell enCore FC Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Honeywell enCore FC Series and is the answer not in the manual?

Honeywell enCore FC Series Specifications

General IconGeneral
BrandHoneywell
ModelenCore FC Series
CategoryDesktop
LanguageEnglish

Related product manuals