MESE
MESE
command/query
The MESE command sets the Module Event Status Enable Register bits.
The MESE register contains a mask value for the bits enabled in the
MESR register. A one in the MESE will enable the corresponding bit in
the MESR, a zero will disable the bit.
The MESE query returns the current setting.
Refer to table 6-2 for information about the Module Event Status Enable
register bits, bit weights, and what each bit masks for the logic analyzer.
Command Syntax:
:SYSTem:MESE
<enable-mask>
where:
-Z
enable mask >
:: = integer from 0 to 255
Example: OUTPUT
XXX;“:SYSTEM:MESE
1”