Command Syntax:
Example:
Query Command:
Returned Format:
Example:
command/query
The HEADER command tells the instrument whether or not to output a
header for query responses. When HEADer is set to ON, query
responses will include the command header.
The HEADer query returns the current state of the HEADer command.
:SYSTem:HEADer
{{ON~1}~{OFFJO}}
OUTPUT XXX;" :SYSTEM:HEADER ON"
:SYSTem:HEADer?
[:SYSTem:HEADer]
{llO}c
NL>
10 DIM Mode$[lOO]
20 OUTPUT XXX;":SYSTEM:HEADER?"
30 ENTER XXX;Mode$
40 PRINT Mode$
50 END
Headers should be turned off when returning values to numeric variables.