Page 60 Setup and Assembly Manual
C.6.2 How to Recover to Factory Defaults
1. Power the device down with the stop button.
Picture 59: Connectors on the WideTEK 25
1. Power connector
and main power
switch
2. Serial connector
3. VGA connector
4. Two foot pedal
connectors
5. Network cable
connector
2. Plug the recovery key into the serial connector (detail 2 in the above picture) at the
connector panel.
3. Power the device up via the start button.
The device will start and, after it has found the recovery key to be present in the serial
port, it will automatically execute the recovery sequence. All viable system data will be
restored and necessary repair steps will be taken without the need of any user interaction.
Note: The recovery sequence can last some minutes. While the recovery sequence
is running, no message will be displayed.
4. When the recovery sequence has finished, the device will power down automatically.
Important: Do not switch off the device at any time during the recovery procedure!
5. Unplug the recovery key after the device has powered down.
6. Power up the device and launch the scan application in your browser.
The IP address of the device will have the factory default value: 192.168.1.50