Inspection and Installation
Copyright ©ITECH Electronic Co., Ltd. 31
1. Confirm that the power switch is in the OFF position and verify that there is
no dangerous voltage on the connection terminals.
2. Remove the output terminals cover of the power system.
3. Loosen the screws of the output terminals and connect the red and black
test cables to the output terminals. Re-tighten the screws.
When maximum current that one test cable can withstand fails to meet the
rated current, use multiple pieces of red and black test cables. For example,
the maximum current is 1,200A, then 4 pieces of 360A red and black cables
are required.
4. Thread the red and black test cables through the output terminals cover of
the power system and install the cover.
5. (Optional) According to the actual situation of DUT, connect the grounding
terminal on the rear panel of the instrument to the DUT to ensure the safe
grounding.
For the location information, see 1.5 Rear Panel Introduction.
6. Connect the other end of the red and black cables to the DUT. The positive
and negative poles must be properly connected and fastened when wiring.
Connecting the DUT (Remote Sensing)
Remote measurement is available for the following scenarios:
When the DUT consumes large current or the wires are too long, there is a
voltage drop on the wires between DUT and output terminals of the power
system.
To maximize measurement accuracy, the power system provides the remote
measurement terminals VS+ and VS- on the rear panel, which can be used to
measure the terminal voltage of the DUT.
When the power system is used for battery testing in actual applications, the
voltage drop of the wire will lead to voltage inconsistency of both ends and
inconsistency of the cutoff voltage of power system and the actual voltage of
battery, resulting in inaccurate measurement.
The connection diagram and steps of remote measurement are as follows:
NOTE
The T7830-350-180 consists of two 15kVA in parallel mode. Connect the Remote
sense terminals of master module. The wiring method is the same as other
models.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ITech IT7800 Series and is the answer not in the manual?

ITech IT7800 Series Specifications

General IconGeneral
BrandITech
ModelIT7800 Series
CategoryAC Power Distribution
LanguageEnglish

Related product manuals