Page #106 background image
Model 2110
5½ Digit Multimeter Reference Manual Section 4:
Theory of operation
2110-901-01 Rev. C/August 2013 4-5
Accuracy calculations
The following information discusses how to calculate accuracy for both DC and AC characteristics.
Calculating DC and AC characteristics accuracy
DC characteristics accuracy is calculated as follows:
Accuracy =
% of reading + % of range
As an example of how to calculate the actual reading limits, assume that you are measuring 5V on
the 10V range. You can compute the reading limit range from one-year DCV accuracy specifications
as follows:
Accuracy =
0.012% of reading + 0.004% of range
0.012% * 5 V + 0.004% * 10 V
0.0006 V + 0.0004 V
0.0010 V
Thus, the actual reading range is 5 V ± 1 mV or from 4.999 V to 5.001 V.
DC current, AC voltage, AC current, and resistance calculations are performed in exactly the same
manner using the pertinent specifications, ranges, and input signal values.

Table of Contents

Other manuals for Keithley 2110

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2110 and is the answer not in the manual?

Keithley 2110 Specifications

General IconGeneral
Model2110
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
ContinuityYes
Diode TestYes
CategoryDigital Multimeter
Display5.5 digit
DC Current Range100 µA to 3 A
AC Current Range100 µA to 3 A
Resistance Range100 MΩ
Frequency Range3 Hz to 300 kHz
Frequency Accuracy0.01% + 1 count
Capacitance Range1 nF to 100 µF
Capacitance Accuracy1% + 5 counts
ConnectivityUSB, GPIB (optional)
Dimensions88 mm
Power Supply100 V to 240 V AC, 50/60 Hz
DC Voltage Accuracy±0.012% + 2 counts