5½ Digit Multimeter Reference Manual Section 5:
2110-901-01 Rev. C/August 2013 5-31
CALCulate:MXB:MMFactor <n>|MIN|MAX
This command defines the m factor for the mX+b calculation.
Parameters
<n> description
Set value to the minimum.
Set value to the maximum.
Query
command description
CALCulate:MXB:MMFactor? MIN
Query the minimum value of m.
CALCulate:MXB:MMFactor? MAX
Query the maximum value of m.
CALCulate:NULL:OFFSet <n>|MIN|MAX
This command stores a null value in the instrument’s Null Register. You must turn on the math operation before
writing to the math register.
Parameters
<n> description
0 to ± 120% of the
highest range
Select the null (relative) value. You can set the null value to any number
between -120 % and 120 % of the highest range.
Set value to the minimum.
Set value to the maximum.
Query
command description
Query the null minimum value for null math function.
Query the null maximum value for null math function.
CALCulate:PERCent:TARGet <n>|MIN|MAX
This command specifies the target value for percent math function.
Parameters
<n> description
Specify the measurement function dependent target value.