5½ Digit Multimeter Reference Manual Section 5:
2110-901-01 Rev. C/August 2013 5-53
[SENSe:]TEMPerature:NTCT:C <n>|MIN|MAX
Use this command to select the C coefficient for the NTCT type.
Parameters
<n> description
Select the C coefficient.
Select the minimum coefficient.
Select the maximum coefficient.
Query
command description
[SENSe:]TEMPerature:NTCT:C?
Query the C coefficient for the NTCT type.
[SENSe:]TEMPerature:NTCT:C? MIN
Query the minimum C coefficient for the NTCT type.
[SENSe:]TEMPerature:NTCT:C? MAX
Query the maximum C coefficient for the NTCT type.
Details
Refer to the following topic in the Temperature measurement (on page 3-17) section for more information about
the NTCT type: NTCT temperature calculation (on page 3-19).
[SENSe:]TEMPerature:RTD:ALPHa <n>|MIN|MAX
Use this command to select the alpha coefficient for the user-defined RTD type.
Parameters
<n> description
Select the alpha coefficient.
Select the minimum coefficient.
Select the maximum coefficient.
Query
command description
[SENSe:]TEMPerature:RTD:ALPHa?
Query the alpha coefficient for the user defined RTD
type.
[SENSe:]TEMPerature:RTD:ALPHa? MIN
Query the minimum alpha coefficient for the user
defined RTD type.
[SENSe:]TEMPerature:RTD:ALPHa? MAX
Query the maximum alpha coefficient for the user
defined RTD type.
Details
Refer to the following topic in the Temperature measurement (on page 3-17) section for more information about
RTD type: RTD coefficients and temperature calculation (on page 3-19).