EasyManuals Logo
Home>Keithley>Multimeter>2110

Keithley 2110 User Manual

Keithley 2110
221 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #153 background imageLoading...
Page #153 background image
Section
5: Remote commands Model 2110 5½ Digit Multimeter
Reference Manual
5-46 2110-901-01 Rev. C/August 2013
[SENSe:]PERiod:APERture <n>|MIN|MAX
Use this command to set the aperture time for period function. For frequency and period measurements, aperture
time is analogous to integration time.
Parameters
<n> description
0.01
4½-digit resolution.
0.1
5½-digit resolution.
1
5½-digit resolution.
Query
command description
[SENSe:]PERiod:APERture?
Query the aperture time.
[SENSe:]PERiod:APERture? MIN
Query the minimum aperture time.
[SENSe:]PERiod:APERture? MAX
Query the maximum aperture time.
Details
Refer to Aperture and resolution for frequency and period measurements (on page 3-45) for more information
about aperture time.
[SENSe:]PERiod:VOLTage:RANGe <n>|MIN|MAX
Use this command to select the voltage range for the period function.
Parameters
<n> description
0 to 750.
Set AC voltage range for the period function.
MINimum
Select the minimum value.
MAXimum
Select the maximum value.
Query
command description
[SENSe:]PERiod:VOLTage:RANGe?
Query the measurement range.
[SENSe:]PERiod:VOLTage:RANGe? MIN
Query the minimum range.
[SENSe:]PERiod:VOLTage:RANGe? MAX
Query the maximum range.

Table of Contents

Other manuals for Keithley 2110

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2110 and is the answer not in the manual?

Keithley 2110 Specifications

General IconGeneral
BrandKeithley
Model2110
CategoryMultimeter
LanguageEnglish

Related product manuals