5½ Digit Multimeter Reference Manual Section 5:
2110-901-01 Rev. C/August 2013 5-21
DISPlay subsystem summary
Command Description
Turn on or off front panel display. Where: 0 = off and 1 = on.
Define ASCII message “a” (up to 16 character
s in length). The message
will display on the secondary display.
Clear the message shown on the front panel display.
Query the display setting. Returns 0 (off) or 1 (on).
SENSe subsystem summary
See the following table for the SENSe command summary. Refer to Power-on settings (on page 1-6)
for information about default values for SENSe commands.
Command Description
Path to FUNCtion, FUNCtion1, and FUNCtion2. FUNCtion1
displays the selected function on the primary display. FUNCtion2
displays the selected function on the secondary display.
Specify one of the following functions to measure (enclosed
function in quotation marks):
Note: DIODe and CONTinuity are not SCPI compliant
parameters.
Query function. Returns function name.
Path to configure DC voltage.
Specify the range: 100 mV, 1 V, 10 V, 100 V, 1000 V, MIN, or
MAX.
Enable or disable auto range. Where: 0 = OFF and 1 = ON.
Specify resolution: <n> = 3.00e-07 to 3.00e-04 of range, MIN, or
MAX. Refer to Integration time and resolution for DC and
resistance measurements (on page 3-41) for more information
about valid values for n.