5½ Digit Multimeter Reference Manual Section 5:
2110-901-01 Rev. C/August 2013 5-23
Command Description
Specify the integration time in number of power line cycles for
the selected function: 0.001, 0.006, 0.02, 0.06, 1, 2, 10, 100,
MIN, or MAX. Refer to
Integration time and resolution for DC and
resistance measurements (on page 3-41) for more information
about valid values for n.
Path to configure fresistance (4-wire
Ω
Specify range: 0 to 100e6, MIN, or MAX.
Query the fresistance range.
Enable or disable auto range. Where, 0 = OFF and 1 = ON.
Specify resolution: <n> = 3.00e-07 to 3.00e-04 of range, MIN, or
MAX. Refer to Integration time and resolution for DC and
resistance measurements (on page 3-41) for more information
about valid values for n.
Query the fresistance resolution.
Specify the integration time in number of power line cycles for
the selected function: 0.001, 0.006, 0.02, 0.1, 1, 2, 10, 100, MIN,
Path to configure frequency.
Path to configure voltage.
Specify range: 0 to 750, MIN, or MAX.
Enable or disable auto range. Where: 0 = OFF and 1 = ON.
Query auto range setting.
Path to configure current.
Specify the aperture time for the selected function: 0.01, 0.1, 1,
MIN, or MAX.
Query the aperture time for frequency function.
Path to configure period.
Specify range: 0 to 750, MIN, or MAX.
Query the period range setting.
Enable or disable auto range. Where: 0 = OFF and 1 = ON.
Query auto range setting.
Path to configure current.
Specify the aperture time for the selected function: 0.01, 0.1, 1,
MIN, or MAX. Default = 100 ms.
Query the aperture time for period function.
Specify units for temperature measurement (C, F, or K)
Query units for temperature measurement.
Path to configure thermocouple.
Specify thermocouple sensor type (B, C, E, J, K, N, R, S, or T).
Query thermocouple sensor type.
Path to thermocouple reference junction.
Specify reference junction type, real or simulated. (REAL or
SIMulated).