A- 2 Specifications
2430 Specifications
Source Specifications
1
Voltage Programming Accuracy (remote sense)
Accuracy (1 Year) Noise
Programming 23°C ±5°C (peak-peak)
Range Resolution ±% rdg. + volt s 0.1Hz – 10Hz
200.000 m V 5 V 0.02% + 600 V 10 V
2.00000 V 50 V 0.02% + 600 V 50 V
20.0000 V 500 V 0.02% + 2.4 m V 500 V
100.000 V 2.5 m V 0.02% + 12 m V 2.5 m V
TEMPERATURE COEFFICIENT (0°–18°C & 28°–50°C): ±(0.15 accuracy specification)/ °C.
MAX. OUTPUT POWER: 110W (four quadrant source or sink operation, DC mode).
SOURCE/SINK LIMITS: 5% over nominal for current and volt age.
VOLTAGE REGULATION: Line: 0.01% of range. Load: 0.01% of range + 100 V.
NOISE 10Hz–1MHz (p-p): 50mV maximum, resistive load.
OVER VOLTAGE PROTECTION: User selectable values, 5% tolerance. Factory default = None.
CURRENT LIMIT: Bipolar current limit (compliance) set with single value. Min. 0.1% of range.
OVERSHOOT:<0.1% typical (full scale step, resistive load, 10mA range).
Current Programming Accuracy (local or remote sense)
Accuracy (1 Year) Noise
Programming 23°C ±5°C (peak-peak)
Range Resolution ±% rdg. +amps 0.1Hz – 10Hz
10.0000 A 500 pA 0.033% + 2 nA 50 pA
100.000 A 5 nA 0.031% + 20 nA 500 pA
1.00000 m A 50 nA 0.034% + 200 nA 5 nA
10.0000 m A 500 nA 0.045% + 2 A 50 nA
100.000 m A 5 A 0.066% + 20 A 500 nA
1.00000 A
2
50 A 0.067% + 400 A 100 A
3.00000 A
2
/ 500 A 0.059% + 2.8 m A 300 A
10.0000 A
3
500 A 0.089% + 5.9 m A 300 A
TEMPERATURE COEFFICIENT (0°–18°C & 28°–50°C): ±(0.15 accuracy specification)/ °C.
MAX. OUTPUT POWER: 110W (four quadrant source or sink operation, DC mode)
2
. 1100W (four quadrant source or sink operation, PULSE mode)
3
.
SOURCE/SINK LIMITS: ±1.05A @ 105V, ±3.15A @ 2 1 V, ±10.5A @ 105V 2.5ms pulse.
CURRENT REGULATION: Line: 0.01% of range. Load: 0.01% of range + 100pA.
V O LTAGE LIMIT:Bipolar volt age limit (compliance) set with single value. Min. 0.1% of range.
OVERSHOOT:<0.1% typical (1mA step, R
L = 10kž, 20V range).
Additional Source Specifications
TRANSIENT RESPONSE TIME: 30 s typical for the output to recover to it s spec. following a step change in load. Resistive load.
COMMAND PROCESSING TIME: Maximum time required for the output to begin to change following the receipt of :SOURce:VOLTage|CURRent
<nrf> command.Autorange On: 10ms. Autorange Off: 7ms.
OUTPUT SETTLING TIME: Time required to reach 0.1% of final value af ter command is processed. 100 s typical. Resistive load. 10 A to 100mA
range, 1V step.
OUTPUT SLEW RATE: 0.25V/ s ±30%, 100V range, 100mA compliance. 0.08V/ s ±30%, 20V range, 100mA compliance.
COMMON MODE VOLTAGE: 250VDC.
REMOTE SENSE: Up to 1V drop per load lead.
COMPLIANCE ACCURACY:For Source mode, add 0.1% of range to base specification. For Sink mode, 10 A to 100mA range, accuracy is ±(0.5%
+ of fset
*
3). For Sink mode, 1A and 3A ranges, accuracy is ±(1.5% + offset
*
3).
OVER TEMPERATURE PROTECTION: Internally sensed temperature overload put s unit in st andby mode.
RANGE CHANGE OVERSHOOT:Overshoot into a fully resistive 100kž load, 10Hz to 1MHz BW, adjacent ranges: (100mV) typical, except 20V/100V
range boundary.
MINIMUM COMPLIANCE VALUE: 0.1% of range.
Section A (pgs 1-5).qxd 11/7/2005 1:35 PM Page 2

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2430 and is the answer not in the manual?

Keithley 2430 Specifications

General IconGeneral
BrandKeithley
Model2430
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals