Specifications A-13
Calculating dBm characteristics accuracy
As an example of how to calculate the actual reading limits for a 13dBm measurement with
a reference impedance of 50, assume an applied signal 0.998815V. The relationship between
voltage and dBm is as follows:
From the previous example on calculating DC characteristics accuracy, it can be shown that
a measurement of 0.998815V on the 1V range has an uncertainty of ±36.9644mV, or
0.998778V to 0.998852V, using one-year specifications.
Expressing 0.998778V as dBm:
and expressing 0.998852V as dBm:
Thus, the actual reading range is 13dBm ±0.00032dBm.
dBm and dB for other voltage inputs can be calculated in exactly the same manner using
pertinent specifications, ranges, and reference impedances.
Calculating dB characteristics accuracy
The relationship between voltage and dB is as follows:
As an example of how to calculate the actual readings limits for dB, with a user-defined
VREF of 10V, you must calculate the voltage accuracy and apply it to the above equation.
To calculate a -60dB measurement, assume 10mV RMS for a VREF of 10V. Using the
100mV range, one-year, 10Hz - 20kHz frequency band, and SLOW rate, the voltage limits are
as follows:
Accuracy = ±[(0.06% of reading) + (0.03% of range)]
±[(0.0006 × 10mV) + (0.0003 × 100mV)]
±[6µV + 30µV]
±36µV
dBm 10
V
2
IN
R
REF
1mW
---------------------------log=
dBm 10
0.998778V()
2
50
1mW
-------------------------------------------------log 12.99968dBm==
dBm
0.998852V()
2
50
1mW
------------------------------------------------- 13.00032dBm==
dB 20
V
IN
V
REF
------------log=

Table of Contents

Other manuals for Keithley 2700

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2700 and is the answer not in the manual?

Keithley 2700 Specifications

General IconGeneral
BrandKeithley
Model2700
CategoryMultimeter
LanguageEnglish

Related product manuals