EasyManuals Logo
Home>Keithley>Multimeter>2700

Keithley 2700 Service Manual

Keithley 2700
174 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #135 background imageLoading...
Page #135 background image
Specifications A-15
Optimizing measurement speed
The configurations listed below assume that the multimeter has had factory setups restored.
DC voltage, DC current, and resistance:
Select 3-1/2 digits, 0.01 PLC, filter OFF, fixed range.
AC voltage and AC current:
Select 3-1/2 digits, 0.01 PLC, filter OFF, fixed range.
Temperature:
Select 3-1/2 digits, 0.01 PLC, filter OFF.
For all functions, turn off the display and autozero, and set the trigger delay to zero. Use the
:SAMPle:COUNt and READ? bus commands.