Calibration Reference B-7
:DATE
(:CALibration:PROTected:DATE)
Purpose To send the calibration date to the instrument.
Format :cal:prot:date <year>, <month>, <day>
Parameter <year> = 1999 to 2098
<month> = 1 to 12
<day> = 1 to 31
Query format :cal:prot:date?
Response <year>, <month>, <day>
Description The :DATE command allows you to store the calibration date in instrument
memory for future reference. You can read back the date from the instru-
ment over the bus by using the :DATE? query or the CALIBRATION
selection in the front panel CAL menu.
NOTE The year, month, and day parameters must be delimited by commas.
Example :CAL:PROT:DATE 1999,12,16 Send cal date (12/16/1999).
:CAL:PROT:DATE? Request cal date.
:NDUE
:CALibration:PROTected:NDUE)
Purpose To send the next calibration due date to the instrument.
Format :cal:prot:ndue <year>, <month>, <day>
Parameter <year> = 1999 to 2098
<month> = 1 to 12
<day> = 1 to 31
Query format :cal:prot:ndue?
Response <year>, <month>, <day>
Description The :NDUE command allows you to store the date when calibration is next
due in instrument memory. You can read back the next due date from the
instrument over the bus by using the :NDUE? query or the front panel CAL
menu.
NOTE The next due date parameters must be delimited by commas.
Example :CAL:PROT:NDUE 2000,12,16 Send due date (12/16/2000).
:CAL:PROT:NDUE? Request due date.

Table of Contents

Other manuals for Keithley 2700

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2700 and is the answer not in the manual?

Keithley 2700 Specifications

General IconGeneral
BrandKeithley
Model2700
CategoryMultimeter
LanguageEnglish

Related product manuals