EasyManuals Logo
Home>Megger>Test Equipment>TRAX

Megger TRAX User Manual

Megger TRAX
58 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #36 background imageLoading...
Page #36 background image
36 TRAX ZP-AJ01E AJ0383DE
5 OPERATING INSTRUCTIONS
If OLTC
Important
Do not use autostop when testing OLTC.
OLTC measurements must be performed
by winding (see global settings).
1] When reading is stable, press
The measurement and the result is captured
and displayed for the actual row/tap.
Measurement will continue on the next table
row (next tap position).
Note Do not stop generating until last tap is tested.
2] Operate tap changer.
A] If a discontinuity occurs (brake before
make), TRAX will automatically stop the
test, discharge the winding and report
fail in the continuity column.
The discontinuity is reported on the row/
tap that the switch transition is going TO,
i.e a transition discountinuity when going
from tap 5 to 6 will be reported on the
row for tap 6.
B] If continuity is OK, TRAX will start meas-
ure resistance for the actual tap on the
active row.
Wait for stable reading and press
TRAX will continue to measure next posi-
tion.
3] Repeat from step 2.
Continue until last tap.
4] Stop measurement (generator) and capture
data when the reading is stable.
5] Reconnect cables and perform test on next
phase.
Winding variation will be displayed when all phases
are measured.
5.6 Demagnetization
Important
Read and comply with the safety instruc-
tions “2 Safety” on page 8.
Always comply with local safety regula-
tions.
1] Press
Demagnetization is recommended before performing
any tests on the transformer and in particular before
excitation current and/or SFRA.
Demagnetization with TRAX is performed by injecting
an alternating and decreasing DC voltage / current to
magnetize the core in two directions. Starting current
is normally selected as about the same as the last per-
formed winding resistance test and should be above
the DC saturation level of the actual winding (typically
1 % of rated winding current). Demagnetization is
normally done on the HV side of the transformer and
on the connection with the lowest excitation current
(the mid-leg for a YN configuration). For configura-
tions with no neutral, demagnetization connection is
terminal-terminal.
The adaptive algorithm for the demagnetization pro-
cess is based on measuring and reducing Vs (voltage
* time). This implies that the voltage measurement R1
MUST be connected during demagnetization.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Megger TRAX and is the answer not in the manual?

Megger TRAX Specifications

General IconGeneral
BrandMegger
ModelTRAX
CategoryTest Equipment
LanguageEnglish

Related product manuals