EasyManuals Logo
Home>Metrohm>Measuring Instruments>912

Metrohm 912 User Manual

Metrohm 912
106 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #100 background imageLoading...
Page #100 background image
8.2 Measured value memory
■■■■■■■■■■■■■■■■■■■■■■
90
■■■■■■■■
912/913/914 Meter
Measuring range Resolution Measuring accuracy
2)
Oxygen 0.0 - +500%
0.00 - +50.00 mg/L
0.1%
0.01 mg/L
in the range 0 - 8 mg/L: +/-0.1 mg/L
in the range 8 - 20 mg/L: 0.15 mg/L
in the range 20 - 50 mg/L: 10%
2)
±1 digit, without sensor error, at reference conditions
3)
To obtain the conductivity, the corresponding value must be multiplied
by the cell constant. The indicated values apply for c = 1/cm.
Display interval of the measurement = 1 s
8.2 Measured value memory
Memory size
10000 measured values, non-volatile memory
10 sensor entries in sensor list
8.3 TFT display
Resolution
320 x 240 pixels (RGB)
Display colors 16.7 millions
Display size 3.5 inches (70.08 x 52.56 mm)
8.4 Interfaces
USB connector
Type A/B mini USB connector (USB 2.0) with the following functions:
Energy supply
Data transmission
with USB cable (6.2151.110)
Printing
with USB Y cable (6.2151.140)