4.7 Menu dialogs
■■■■■■■■■■■■■■■■■■■■■■
48
■■■■■■■■
912/913/914 Meter
■ drift-dependent
The measured value will be applied when the value is stable according to the
drift criterion.
The drift thresholds are preset and cannot be changed:
– pH measurement: 0.028 pH/min
– Potential measurement U/mV: 1.875 mV/min
– Temperature measurement T/°C: 0.974 °C/min
– Oxygen measurement DO 0.24 mg/L/min
For the conductivity, various drift thresholds are stored (depending on the
measuring range):
– 0.005 mS/cm/min in the measuring range up to 16 µS/cm
– 0.5 mS/cm/min in the measuring range from 16 µS/cm to 1 mS/cm
– 10 mS/cm/min in the measuring range from 1 mS/cm
For instruments with two measuring channels, the primary measuring chan-
nel for the fulfillment of the drift criterion has to be selected.
– Primary channel
Selection dialog for selecting the measuring channel that has to fulfill
the drift criterion.
pH/mV
pH/mV IS
Cond.
DO IS
Output date/time
Selection dialog for selecting whether a time stamp is to be assigned to a mea-
sured value.
■ on: Measured values contain a time stamp on the report.
■ off: Measured values do not contain a time stamp on the report.
Default value: off
Output headers
Selection dialog for selecting how the headers are output.
■ once
■ always: is the default value
■ off
Calibration data
Selection dialog for selecting whether the main Calibration data is assigned for
the output of the measured values.
■ on
■ off: is the default value
4.7.5 Sensors
The Sensors menu dialog is shown below with the structure and the
description.