4.8 pH measurement
■■■■■■■■■■■■■■■■■■■■■■
60
■■■■■■■■
912/913/914 Meter
4.8.2 Measurement
NOTICE
Measured value criteria
You can set the various criteria for defining the measured value determi-
nation as follows (see chapter 4.7.4, page 46):
1
Selecting the printout criterion
■ If the measured value found is to be directly printed out as a mea-
sured value report, then you have to set the required printout cri-
terion (see chapter 4.5.3, page 31).
2
Selecting the measured value criterion
■ This criterion defines the conditions as to when the measured
value is saved on the instrument and/or printed out.
3
Carrying out the measurement
■ Rinse the sensor with water and immerse it in the sample.
■ Select the Print/save measured value button with the
keys.
■ Trigger printing and/or saving of the measured value with the
key.