Accutorr V Operating Instructions 0070-10-0699-02 6 - 3
Appendix Specifications
6.2 Specifications
6.2.1 Systolic Pressure Readout
6.2.2 Diastolic Pressure Readout
6.2.3 Mean Pressure Readout
6.2.4 NIBP Measurement Cycle Time
Less than 40 seconds average at 80 BPM with 180mmHg pump up pressure, without retries,
motion artifact or arrhythmia with standard adult cuff on a healthy individual. Cycle time is
affected by arm size and wrapping technique.
Number of Digits: 3
Accuracy: Mean error: within ±5 mmHg
Standard deviation: within ±8 mmHg
Range: Adult Mode: 55 – 235 mmHg
Pediatric Mode: 55 – 160 mmHg
Neonatal Mode: 45 – 120 mmHg
Measurements outside of the stated ranges are not guaranteed to be accurate.
Number of Digits: 3
Accuracy: Mean error: within ±5 mmHg
Standard deviation: within ±8 mmHg
Range: Adult Mode: 30 – 200 mmHg
Pediatric Mode: 30 – 150 mmHg
Neonatal Mode: 20 – 100 mmHg
Measurements outside of the stated ranges are not guaranteed to be accurate.
Number of Digits: 3
Accuracy: Mean error: <±5 mmHg
Standard deviation: <±8 mmHg
Range: 30 – 235 mmHg for Adult
30 – 160 mmHg for Pediatric
20 – 120 mmHg for Neonates
Measurements outside of the stated ranges are not guaranteed to be accurate.