Accutorr V Operating Instructions 0070-10-0699-02 6 - 11
Appendix Recovery from Power Loss
6.3 Recovery from Power Loss
6.3.1 Alarm Restoration from Power Loss
If the Accutorr V is turned off or loses power for less than 60 seconds, the alarm settings in
use at that time are restored when it is turned on.
If the Accutorr V is turned off or loses power for more than 60 seconds, the default power-on
configuration is restored (see Section 3.16.6, “Setting a Default Power-on Configuration”).
6.3.2 Data Logging after Power Loss
The patient database will be restored following a power loss of any duration.