Specifications Appendix
6 - 8 0070-10-0699-02 Accutorr V Operating Instructions
6.2.10 DPM Performance Specifications
SpO
2
6.2.11 Battery
Range: 0 – 100%
Display Resolution: 1%
Averaging Time 8 seconds
Typical response Time 15 seconds
Alarm Condition delay plus
alarm generation
<1 second
Calibration: Factory Calibrated to Functional Saturation
Accuracy: Adult and Pediatric: 70% – 100% ±2 digits
Adult and Pediatric: 0% – 69%, Undefined
Neonate: 70% – 100% ±3 digits
Neonate: 0% – 69%, Undefined
Pulse Rate
Pulse Rate Range 20 – 254 BPM
Display Resolution: 1 BPM
Accuracy: Adult / Pediatric / Neonate: 20 – 254 BPM ±3 digits
Update Rate: 1 second
Wavelength of light emitted by
the sensors
red: 660 nm
infrared 905 nm
Battery Type: Lithium
ion
Number of Batteries: 1
Battery Voltage: 11.1 VDC nominal
Battery Capacity: 4.6 Amp-Hour