Accutorr V Operating Instructions 0070-10-0699-02 3 - 55
Operation Status and Error Codes
3.17.10 System Alarm Messages
3.17.11 Prompt Messages
ALARM MESSAGE A B LEVEL CAUSE ACTION
REAL CLOCK NEED SET No No L The system time is
incorrect.
Reset the system time and
then restart the monitor.
REAL CLK NOT EXIST No No L There is no button
battery, or the battery
power is depleted.
Install a button battery or
replace with a new one.
NET INIT ERR (G.) No No L The monitor cannot be
connected to the
network due to network
problem.
Contact service personnel
for repair.
NET INIT ERR (RAM) No No L
NET INIT ERR (REG) No No L
NET INIT ERR (MII) No No L
NET INIT ERR (LOOP) No No L
NET ERR (RUN 1) No No L
NET ERR (RUN 2) No No L
NET ERR (RUN 3) No No L
12V TOO HIGH No No L A problem occurs to the
system’s power supply.
If this alarm message
occurs frequently, contact
service personnel for
repair.
12V TOO LOW No No L
BATTERY TOO LOW No No H The battery voltage is
too low.
Connect the monitor to AC
mains to charge the
battery.
NET ERROR Yes No L The monitor is not
connected to the
network.
Check the network links.
The “Afield indicates whether all alarm indications can be cleared or not, and the “B” field indicates whether all alarm
indications except the alarm message can be cleared or not. The alarm levels are H = high and L = low.
PROMPT MESSAGE CAUSE ACTION
PULSE SEARCH The SpO
2
module is searching the pulse. Wait till the end of the search.
REC INITIALIZING The recorder is initializing Wait until initialization complete before
printing.
NIBP MEASURING... NIBP measurement in process Wait until measurement complete or
press (25) to interrupt a
measurement and deflate the cuff.
PRESS START NIBP Interval NIBP measurement has been
stopped by pressing (25).
Press
(24) to restart NIBP
interval mode measurements.
PNEUM TESTING... Pneumatic testing in process. Wait until testing complete.
MEASUREMENT
COMPLETE
A measurement has finished. None
NIBP RETRY NIBP measurement failed and is retrying. Wait until measurement complete or
press (25) to interrupt a
measurement and deflate the cuff.
NIBP RETRY PUMP HIGHER NIBP measurement failed and is retrying
with a higher pressure.
BARCODE FAILED The barcode reader failed to read the
patient’s barcode.
Use another barcode reader. If failure
continues, contact service personnel.