Alarms Operation
3 - 20 0070-10-0699-02 Accutorr V Operating Instructions
3.8.3 Pausing and Silencing Alarms
When an NIBP alarm exists:
Press (29) for less than two (2) seconds to pause the alarm tone for two (2)
minutes. The alarm tone returns if the next measurement value violates the selected
limits.
NOTE: The Accutorr V counts down time remaining for the alarm
pause in the Trend display.
Press (29) for two (2) seconds or more to silence the alarm. The alarm tone
returns after the next measurement value that violates the selected limits.
When in the paused state:
Press (29) for less than two (2) seconds to return to the normal state.
Press (29) for two (2) seconds or more to silence the alarm. The alarm tone
returns after the next measurement value that violates the selected limits.
When in the silenced state:
Press (29) for less than two (2) seconds to enter the paused state.
Press (29) for two (2) seconds or more to return to the normal state.