Accutorr V Operating Instructions 0070-10-0699-02 3 - 41
Operation Battery Operation
3.15 Battery Operation
When the Accutorr V is powered from the battery and switched on, the Battery status
indicator (15) illuminates, and the AC power indicator (14) remains dark.
When the battery charge is low, but not below the cutoff voltage, the battery indicator
flashes. When the LED begins to flash, at least 5 minutes of low battery warning time remain.
Battery run time for the Accutorr V is approximately 540 minutes for a new Lithium ion
battery, fully charged, at an ambient temperature of 25
°
C, with one automatic NIBP
measurement taken every 15 minutes, continuous SpO
2
measurement, and the recorder not
in use.
The Accutorr V automatically recharges the battery, when required, if the monitor is plugged
into an AC receptacle. When the unit is plugged into an AC receptacle, the Battery status
indicator (15) is always illuminated. Maximum battery recharge time is 4.5 hours for Lithium
ion with the Accutorr V in standby mode or off, and 6 hours in normal running mode (not in
Standby mode).
NOTE: Replace the Lithium ion battery with new battery if its run
time is diminished.