Accutorr V Operating Instructions 0070-10-0699-02 3 - 39
Operation Recorder
3.13 Recorder
The Accutorr V provides a permanent record of patient data using the recorder.
• To print the data on the current display (5 lines of data), press (33) for less than 2
seconds (1 beep tone).
• To print all the trend data for the current patient (up to 11 lines of data per block.), press
(33) for more than 2 seconds (1 beep tone) in trend data review mode.
• To stop printing while in progress, press (33) (1 beep tone).
FIGURE 3-24 Example Recorder Strip of Data on the Current Display
• Whenever SpO
2
is active, it is used to measure Pulse Rate.
• If data is invalid or not available for any given parameter, “—” prints under that
parameter.
• If an NIBP measurement was interrupted, “XXX” is printed under Sys, Dia, and Map.
• Parenthesis ( ) indicate a parameter value that violated alarm limits.
Time Sys (mmHg) Dia (mmHg) Map (mmHg) SpO2 (%) PR (BPM) Temp (°F)
12 – 10 11:19 XXX XXX XXX 97.2
Trend Table 12 – 10 11:19 — — — 94 80 96.8
(ALL) 12 – 10 11:18 — — — (85) (122) 95.7
12 – 10 11:18 — — — 92 110 93.4
12 – 10 11:17 120 80 93 — 72 83.5
PATIENT ID: 20081210095209 Record Time: 12-10-2008 11:19:29 Mindray DS USA, Inc.
Parenthesis print around measurements that
caused an alarm violation.
When no information is available for a
particular parameter, a dash prints.
An interrupted NIBP measurement prints as
XXX for Sys, Dia, and Map.
If the temperature is shown in every time slot, the
temperature parameter is in continuous mode.