Status and Error Codes Operation
3 - 54 0070-10-0699-02 Accutorr V Operating Instructions
3.17.9 Recorder Module Alarm Messages
ALARM MESSAGE A B LEVEL CAUSE ACTION
RECORDER INIT ERR N Yes No L An error occurs during
the recorder
initialization.
Contact the hospital’s
engineers or Customer
Service.
REC SELFTEST ERR Yes No L An error might occur to
the RAM, ROM and CPU
watchdog.
Restart the recorder. If the
error remains, contact
service personnel for
repair.
RECORDER HEAD HOT No No L The thermal head of the
recorder is too hot.
Resume the recording after
the recorder cools down
completely. If the problem
still exists, contact service
personnel for repair.
REC HEAD WRONG
POS.
Yes Yes L The thermal head of the
recorder is in wrong
position.
Restore the control lever of
the recorder to its previous
position.
REC OUT OF PAPER Yes Yes L The recorder paper is
used up.
Replace with a new paper
roll.
RECORDER PAPER JAM No No L The recorder has
recorded data on paper
for 30m long or more.
Place the recorder
correctly and try again.
RECORDER COMM ERR Yes No L Error in recorder
communication.
Clear recording tasks and
restart the monitor. If the
problem remains, contact
service personnel for
repair.
TOO MANY REC
TASKS
No No L Too many alarm events
occur at the same time.
RECORDER PAPER W.P. Yes Yes L The paper roll of the
recorder is not placed in
the correct position.
Place the paper roll
correctly.
REC S. COMM ERR Yes No L Error in recorder
communication.
Clear recording tasks and
restart the monitor. If the
problem remains, contact
service personnel for
repair.
REC NOT AVAILABLE No No L Error in the recorder
work mode.
The “A” field indicates whether all alarm indications can be cleared or not, and the “B” field indicates whether all alarm
indications except the alarm message can be cleared or not. The alarm levels are H = high and L = low.