Patient Setup Operation
3 - 8 0070-10-0699-02 Accutorr V Operating Instructions
To select the Patient Size, press (23). Three choices are available: Adult, Pediatric,
and Neonate. The patient size changes with each key press.
The Patient size indicator (21) illuminates to indicate the selected size as shown in
FIGURE 3-6. The factory default Patient Size setting is Adult.
NOTE: Selecting Neonate patient size changes Temperature site
(31) to Axiliary. The temperature site will default to oral
when Adult patient size is selected.
3.5.3 Setting Initial Cuff Inflation Pressure
The initial cuff inflation pressure depends on the Patient Size setting. The initial cuff inflation
pressures are listed in the following table.
NOTE: The default patient size and initial cuff inflation pressure can
be customized.
To modify the initial cuff inflation pressure:
1. Press (12) to display the SYSTEM SETUP dialog as shown in FIGURE 3-7.
2. Press (16) or (19) to highlight MAINTENANCE.
FIGURE 3-7 SYSTEM SETUP Dialog
3. Once MAINTENANCE is highlighted, press (18) to display the MAINTENANCE
dialog as shown in FIGURE 3-8.
PATIENT SIZE
SETTING
FACTORY DEFAULT
INITIAL CUFF
INFLATION VALUES
LOWEST
SELECTABLE
PRESSURE
HIGHEST
SELECTABLE
PRESSURE INCREMENT
Adult 180 mmHg 100 mmHg 280 mmHg 5 mmHg
Pediatric 140 mmHg 60 mmHg 180 mmHg 5 mmHg
Neonate 100 mmHg 40 mmHg 120 mmHg 5 mmHg