EasyManuals Logo
Home>Mindray>Medical Equipment>Accutorr V

Mindray Accutorr V User Manual

Mindray Accutorr V
130 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #73 background imageLoading...
Page #73 background image
Creating a User Configuration Operation
3 - 44 0070-10-0699-02 Accutorr V Operating Instructions
2. Press (16) or (19) to highlight the LANGUAGE selection field.
3. Once the selection field is highlighted, press (18) select it.
4. Press (16) or (19) to highlight a language.
5. Once the selection is highlighted, press (18) select it.
NOTE: The new language becomes effective once the Accutorr V is
restarted.
3.16.3 Turning Alarm Tones Off
To turn off the alarm tones, in the USER MAINTENANCE dialog do the following:
1. Follow steps 1 to 18 above to access the USER MAINTENANCE dialog as shown in
FIGURE 3-28.
2. Press (16) or (19) to highlight the MIN ALARM VOL selection field.
3. Once the selection field is highlighted, press (18) select it.
4. Press (16) or (19) to change the value to 0.
5. Once 0 is displayed, press to set it.
6. Press (12) again to exit to the SYSTEM SETUP dialog.
7. Press (16) or (19) to highlight COMMON SETUP.
8. Once the COMMON SETUP has been highlighted, press (18) to display the
COMMON SETUP dialog as shown in FIGURE 3-14.
9. In steps 1 through 4 of Section 3.10.1, set the ALARM VOL to 0.
3.16.4 SpO
2
Sensor Off
When SPO2 SENSOR OFF is set to OFF, the Accutorr V disables all alarm indications related
to the SpO
2
sensor off alarm.
To set the SpO
2
sensor alarm:
1. Follow steps 1 to 18 above to access the USER MAINTENANCE dialog as shown in
FIGURE 3-28.
2. Press (16) or (19) to highlight the SPO2 SENSOR OFF selection field.
3. Once the selection field is highlighted, press (18) to select it.
4. Press (16) or (19) to change it to OFF.
NOTE: The SPO2 SENSOR OFF selections are HIGH priority, LOW
priority, and OFF.
5. Once OFF is displayed, press to set it.
3.16.5 Saving a user configuration
To save a newly created configuration:
1. Press (16) or (19) to highlight SAVE USER CONFIG.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Accutorr V and is the answer not in the manual?

Mindray Accutorr V Specifications

General IconGeneral
BrandMindray
ModelAccutorr V
CategoryMedical Equipment
LanguageEnglish

Related product manuals