SpO2 Screens SpO2 Monitoring
6 - 2 0070- 0-0704-02 Passport V Operating Instructions
6.2 SpO
2
Screens
6.2.1 Numeric Tile: SpO
2
The SpO
2
numeric tile displays the following:
•SPO
2
& PR label
•SPO
2
& PR Value
•SPO
2
/PR unit
•SpO
2
High and Low Alarm Limits
• Sat-seconds Icon (Nellcor)
• Desat alarm limit
FIGURE 6-1 SpO
2
Numeric Tile
The display rules of Sat-Seconds Icon:
• If Sat-Seconds is set to 10 seconds, 25 seconds, 50 seconds, or 100 seconds, the Sat-
Seconds icon will be displayed.
• The Sat-Seconds icon displays and fills when the low SpO
2
limit is violated. When the set
amount of time has been reached, the SpO
2
alarm will sound.
• The Sat-Seconds icon displays and fills regardless of the “NIBP Simultaneous” setting.
6.2.2 Waveform: SpO
2
• The SpO
2
waveform tile displays the following: Pleth waveform
FIGURE 6-2 SpO
2
Pleth Waveform
NOTE: Wave speed: The Passport V does not provide a separate
Pleth speed setting. ECG speed setting is valid for both Pleth
and ECG waveforms. After ECG waveform speed is
changed, the Pleth shall be refreshed as the speed setting.
NOTE: Scale: The amplitude of the Pleth wave is self-adjusted
according to the height of the waveform area.
High Alarm Limit
Low Alarm Limit
SpO
2
Label
PR Value
SpO
2
Measured Value
Sat-Seconds Icon
Desaturation Limit
SpO
2
Unit
PR Unit
PR Label