Passport V Operating Instructions 0070- 0-0704-02 19 - 25
Appendix Performance Characteristics
19.8.9 Temperature Performance Characteristics
The Passport V is capable of providing temperature measurements from one channel for
400 series probe.
Measuring range: 0 - 50°C (32 -122°F)
Resolution: 0.1°C (0.1°F)
Scale: °C, °F
Tabular Trend: Yes
Graphic Trend: Yes
Clinical Common Range Default of Graphic Trend: 32 - 43°C (90 - 110°F)
Quantitative Influence of
Interference Gas
Interference gas Concentration
Level (%)
Quantitative
Influence
When measuring
under 0~40mmHg,
additional error
introduced due to
gas interference
N
2
O 60 ±1mmHg
HAL 4 ±1mmHg
SEV 5 ±1mmHg
ISO 5 ±1mmHg
ENF 5 ±1mmHg
DES 15 ±2mmHg
Helium Unspecified
Xenon Unspecified
Start Time: 30 s (typical) 60s (maximum)
Warm-up Time: 1 min
Flowrate: Options are 70ml/min, 100ml/min, 120ml/min and
150ml/min.
Flowrate Accuracy: ±15% or ±15ml/min, whichever is greater
0