Passport V Operating Instructions 0070- 0-0704-02 9 - 1
9.0
IBP Monitoring (optional)
9.1 IBP Description
Intended patient types: adult, pediatric, neonatal.
Invasive Blood Pressure (IBP) is a direct measurement of the patient's arterial or venous blood
pressure. IBP utilizes a catheter that is inserted directly into a vein, artery or other pressure
access areas, and is connected to a transducer for interpretation of Systolic (Sys), Diastolic
(Dia), and Mean blood pressures. Up to two Invasive Blood Pressure channels are available.
Invasive Blood Pressure measurement is an option for the Passport V.
The Passport V supports the measurements of two invasive blood pressures. It provides
systolic pressure, diastolic pressure and mean pressure values.
When measuring Art/UA/LV/PA pressure, the channel can serve as an HR source.
When displaying a pulsatile pressure, the format shall be SYS/DIA (MEAN)
When displaying non-pulsatile pressure, the format shall be MEAN (SYS/DIA)
When displaying ICP, the format shall be MEAN CePP.
9.1.1 Pressure Labels
IBP1, IBP2 waveforms may be labeled according to the site of insertion. The labels are
identified in the following table:
LABEL DEFINITION
Art Arterial Blood Pressure
UA Umbilical Artery Pressure
LV Left Ventricular Pressure
PA Pulmonary Artery Pressure
0

Table of Contents

Other manuals for Mindray Passport V

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Passport V and is the answer not in the manual?

Mindray Passport V Specifications

General IconGeneral
BrandMindray
ModelPassport V
CategoryHealth Care Products
LanguageEnglish