Prompt Messages Messages
14 - 16 0070- 0-0704-02 Passport V Operating Instructions
14.3.8 GAS Prompt Message
Caused by Measurement
Error
Measurement Error Repeat calibration
procedure. If problem
persists, contact
Technical Support.
Oridion
Caused by no stable gas flow No stable gas flow Check the gas
connections or gas
concentration. Repeat
calibration procedure.
If problem persists,
contact Technical
Support.
Oridion
ALARM MESSAGE REASON SOLUTION
GM: Agent Uncalibrated Appears after an unsuccessful
calibration attempt of the agent
sensor. The numeric data for all
agents will appear as XXX, and
the agent waveform will be a
flatline.
Ensure proper gas mixture is
attached tightly and regulator
is on. Repeat calibration
procedure. If problem
persists, contact Technical
Support.
GM: CO
2
Uncalibrated Appears after an unsuccessful
calibration attempt of the CO
2
sensor. The numeric data for
CO
2
will appear as XXX, and
the CO
2
waveform will be a
flatline.
Ensure proper gas mixture is
attached tightly and regulator
is on. Repeat calibration
procedure. If problem persists,
contact Mindray
Technical Support.
GM: N
2
O Uncalibrated Appears after an unsuccessful
calibration attempt of the N
2
O
sensor. The numeric data for
N
2
O will appear as XXX, and
the N
2
O waveform will be a
flatline.
Ensure proper gas mixture is
attached tightly and regulator
is on. Repeat calibration
procedure. If problem persists,
contact Mindray
Technical Support.
GM: O
2
Uncalibrated Appears after an unsuccessful
calibration attempt of the O
2
sensor. The numeric data for O
2
will appear as XXX, and the O
2
waveform will be a
flatline.
Ensure proper gas mixture is
attached tightly and regulator
is on. Repeat calibration
procedure. If problem persists,
contact Mindray
Technical Support.
GM: Agent Zero Error Appears when the system has
been unable to successfully zero
the anesthetic agent sensor.
Manually start zeroing the
system again. If problem
persists, contact Mindray
Technical Support.
GM: CO
2
Zero Error Appears when the system has
been unable to successfully zero
the CO
2
sensor.
Manually start zeroing the
system again. If problem
persists, contact Mindray
Technical Support.
GM: N
2
O Zero Error Appears when the system has
been unable to successfully zero
the N
2
O sensor.
Manually start zeroing the
system again. If problem
persists, contact Mindray
Technical Support.
0