Passport V Operating Instructions 0070- 0-0704-02 14 - 5
Messages Technical Alarm Messages
14.2.3 TEMP Alert Message
14.2.4 SpO
2
Alert Message
14.2.5 NIBP Alert Message
Resp Disturbed 3 Patient Movement
External Interference.
Check patient
For impedance - Check
Electrode Contacts /
reposition electrodes/cable
Check environment for source
of interference.
Resp Rate Overrange 3 Respiration Outside
of measureable range
Check patient.
ALARM MESSAGE PRIORITY REASON SOLUTION
Temp Overrange 3 Temperature Outside
of measureable range
Check patient,
Check sensor
ALARM MESSAGE PRIORITY REASON SOLUTION
SpO
2
: Sensor Off 3 Sensor Off Check Patient, Apply Sensor
Note that the alarm is only
active after the sensor has
been on a patient.
SpO
2
: Overrange 3 SpO
2
outside of
measurable range
Check patient
SpO
2
: PR Overrange 3 PR outside of
measurable range
Check patient
ALARM MESSAGE PRIORITY REASON SOLUTION
NIBP: Systolic Overrange 3 NIBP systolic pressure
exceeds the
measurement range.
Check Patient.
Ensure that the measured
pressure label is within the
measurement range.
NIBP: Diastolic Overrange 3 NIBP diastolic
pressure exceeds the
measurement range.
Check Patient.
Ensure that the measured
pressure label is within the
measurement range.
NIBP: Mean Overrange 3 NIBP mean pressure
exceeds the
measurement range.
Check Patient.
Ensure that the measured
pressure label is within the
measurement range.