Care and Cleaning of 3- and 5-lead ECG Cables and Lead wires User Maintenance
17 - 8 0070- 0-0704-02 Passport V Operating Instructions
17.9 Care and Cleaning of 3- and 5-lead ECG Cables
and Lead wires
Recommended cleaning method of ECG cables and lead wires is a cloth wipe using ordinary
alcohol-free hand soap or USP green soap tincture. When disinfection is required, a cloth
wipe using disinfectants such as isopropyl alcohol, chlorine bleach in water (1:10 mixture) or
2% Glutaraldehyde solution (i.e., Cidex) is recommended. After cleaning, the ECG cables
and lead wires should be wiped with water using a clean damp cloth then dried with a clean
dry cloth. The ECG cables and lead wires must be allowed to dry thoroughly before use.
CAUTION: To avoid permanent damage, do not expose metal
components (pins, sockets, snaps) to disinfectants, soaps or
chemicals.
NOTE: ECG cables and lead wires must never be immersed, soaked
in any fluids, and they should not be cleaned with harsh
chemicals such as acetone or non-diluted bleach.
NOTE: Do not autoclave, radiation or steam sterilize ECG cables or
lead wires.
NOTE: Extended exposure to Ethylene Oxide gas may shorten life
of the ECG cables and lead wires, leading to poor signal
quality.
0

Table of Contents

Other manuals for Mindray Passport V

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Passport V and is the answer not in the manual?

Mindray Passport V Specifications

General IconGeneral
BrandMindray
ModelPassport V
CategoryHealth Care Products
LanguageEnglish