Technical Alarm Messages Messages
14 - 6 0070- 0-0704-02 Passport V Operating Instructions
14.2.6 IBP Alert Message
14.2.7 CO
2
Alert Message
14.2.8 GAS Alert Message
ALARM MESSAGE PRIORITY REASON SOLUTION
<IBP Label>: Systolic
Overrange
3 IBP systolic pressure
exceeds the
measurement range.
Ensure that the measured
pressure label is within the
measurement range.
<IBP Label>: Diastolic
Overrange
3 IBP diastolic pressure
exceeds the
measurement range.
Ensure that the measured
pressure label is within the
measurement range.
<IBP Label>: Mean
Overrange
3 IBP mean pressure
exceeds the
measurement range.
Ensure that the measured
pressure label is within the
measurement range.
ALARM MESSAGE PRIORITY REASON SOLUTION
CO
2
: ETCO
2
Overrange 3 Et CO
2
outside of
measurable range
Check patient
CO
2
: InspCO
2
Overrange 3 Insp CO
2
outside of
measureable range
Check patient
CO
2
: Resp Rate
Overrange
3 The measured RR
value exceeds the
measurement range.
Check Patient.
Check Airway.
ALARM MESSAGE PRIORITY REASON SOLUTION
GM: ETCO
2
Overrange 3 The measured
ExpCO
2
value
exceeds the
measurement range.
Check Patient.
Check Airway.
GM: InspCO
2
Overrange 3 The measured InsCO
2
value exceeds the
measurement range.
Check Patient.
Check Airway.
GM: Resp Rate
Overrange
3 The measured RR
value exceeds the
measurement range.
Check Patient.
Check Airway.
GM: Disconnected 2 Appears when the
Passport V cannot
detect signals being
sent by the Gas
Module.
Ensure Gas Module is turned
on and interface cable is
properly connected. If
problem persists, contact
Technical Support.