Passport V Operating Instructions 0070- 0-0704-02 19 - 29
Appendix Communication Protocols
19.10 Communication Protocols
DIAP Communication Protocol
The Passport V supports the proprietary DIAP (0070-00-0307).
Gas Module
The Passport V supports the Gas Module Communication Protocol (0084-00-0025).
Alarm Delay Time to Remote Equipment
The alarm delay time from the Passport V to remote equipment is 2 seconds, measured at
the Passport V signal output connector.
19.11 Data Storage
Monitor Configuration Data Storage and Transfer
Storage
The monitor has the ability to store, in non-volatile memory, the user-selectable configuration.
There is one (1) configuration available for each patient size and the following information is
saved:
Alarm Values
Display Configurations
NIBP Interval
Parameter Settings, Scaling
Trace Speed
Printer Settings
Trend Configurations
Patient Data Storage and Transfer
Storage
The current patient information and demographics is stored in non-volatile memory.
0

Table of Contents

Other manuals for Mindray Passport V

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Passport V and is the answer not in the manual?

Mindray Passport V Specifications

General IconGeneral
BrandMindray
ModelPassport V
CategoryHealth Care Products
LanguageEnglish