Performance Characteristics Appendix
19 - 16 0070- 0-0704-02 Passport V Operating Instructions
NIBP Start Pressure Settings and Ranges
The Start Pressure is adjustable throughout the following ranges and is set to the following
defaults:
NIBP Measurement Cycle
The Maximum Measurement Cycle Duration is 180 sec for Adult and Pediatric patients
(Reference: EN60601-2-30, 1995).
The Maximum Measurement Cycle Duration is 90 sec for Neonatal patients (Reference:
EN60601-2-30, 1995).
Pulse Rate Range for NIBP Measurement
19.8.5 Masimo SET SpO
2
Performance Requirements
Sensor Compatibility
LNOP
®
and LNCS
®
Series
Measured Parameters
SpO
2
and PR
SpO
2
Measurement Range
From 1% to 100%
SpO
2
Accuracy
A functional tester or patient simulator not used to assess the accuracy of the pulse oximeter
probe or a pulse oximeter monitor.
PATIENT SIZE
USER-SELECTABLE START
PRESSURE RANGE INCREMENT SIZE
DEFAULT
SETTING
Adult Mode 100-280 mmHg 5 mmHg 180 mmHg
Pediatric Mode 60-180 mmHg 5 mmHg 140 mmHg
Neonatal Mode 40-120 mmHg 5 mmHg 100 mmHg
PATIENT SIZE RANGE ACCURACY
Adult Mode 35 - 245 bpm The greater of ± 3 bpm or ± 3%
Pediatric Mode 35 - 245 bpm The greater of ± 3 bpm or ± 3%
Neonatal Mode 70 - 245 bpm The greater of ± 3 bpm or ± 3%