NOTE: The watertrap has a filter to prevent bacterium, water, and
secretions from entering the module. After long-term use,
dust or other substances may compromise the performance
of the filter or even block the airway. In this case, replace
the watertrap. It is recommended to replace the watertrap
once every month, or when the watertrap is found leaky,
damaged, or contaminated.
FIGURE 10-5 CO
2
Setup Menu (DPM Sidestream Only)
CO
2
Setup Menu (DPM Sidestream Only)
MENU ITEM SELECTIONS COMMENTS
Previous Screen — Select to return to the previous menu.
Humidity
Compensation
On
Off (default)
Select On for Body Temperature Pressure
Saturation (BTPS).
Select Off for Ambient Temperature and Pressure
Dry (ATPD).
Note: If using the Oridion Medical
miniMediCO
2
module, the humidity
compensation is always on.
O
2
Compensation
0 (default) to 100% Select the O
2
level used for O
2
compensation.
Note: The sum of compensation values input for
oxygen, nitrous oxide and desflurane must not
be greater than 100%.