Passport V Operating Instructions 0070- 0-0704-02 10 - 5
CO
2
Monitoring (optional) CO
2
Menu
10.3.1 CO
2
Setup Menu (DPM Sidestream Only)
To display the CO
2
Setup Menu:
1. On the front panel:
Press the Normal Screen key to return to the normal screen.
On the normal screen:
Select Parameters CO
2
CO
2
Setup
or
Select the CO
2
tile CO
2
Setup.
NOTE: To extend the lifetime of the watertrap and module,
disconnect the watertrap and set the operating mode to
Standby mode when CO
2
monitoring is not required.
NOTE: The watertrap collects water drops condensed in the
sampling line and therefore prevents them from entering
the module. It is important to periodically drain the
watertrap to avoid water blocking the airway.
Selection Gas Module 3
Status
Internal CO
2
Module Status
Active CO
2
Source
Gas Module OFF OFF None
Gas Module OFF ON Internal CO
2
Module
Gas Module ON OFF Gas Module
Gas Module ON ON Gas Module
Internal CO
2
Module
OFF OFF None
Internal CO
2
Module
OFF ON Internal CO
2
Module
Internal CO
2
Module
ON OFF Gas Module
Internal CO
2
Module
ON ON Internal CO
2
Module
Note: ON/OFF Status of the Module:
If the Gas Module is connected to a serial port configured to ‘Gas Module’, and
the power is switched on, the module is considered to be ON. Otherwise it is
considered to be OFF.
If the Oridion CO
2
Module is installed and a Filterline is connected, the module
is considered to be ON. Otherwise it is considered to be OFF.
If the DPM CO
2
Module and water trap are installed, the module is considered
to be ON, Otherwise it is considered to be OFF.
Flow Rate 70 ml/min
100 ml/min (default)
120 ml/min
150 ml/min
Applicable to DPM Sidestream CO
2
module only.
Note: For Oridion Microstream CO
2
, the
flow rate is 50ml/min. It is not user
adjustable.
CO
2
Menu
MENU ITEM SELECTIONS COMMENTS
0

Table of Contents

Other manuals for Mindray Passport V

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Passport V and is the answer not in the manual?

Mindray Passport V Specifications

General IconGeneral
BrandMindray
ModelPassport V
CategoryHealth Care Products
LanguageEnglish