Passport V Operating Instructions 0070- 0-0704-02 19 - 3
Appendix Power Specifications
19.2 Power Specifications
Power Supply Type
AC power, battery
Whole-system consumption:
average consumption of less than 50 W and maximum of 70 W.
AC Mains Power Source
AC power: 100 - 240VAC (±10%)
Current: 1.2 - 0.6 Amps
AC input power frequency: 50 Hz/60 Hz
Internal User-Removable Battery
Lithium-ion battery (M05-010001-06, Mindray)
Number of batteries: 2 (maximum), one battery operation supported
Lithium Battery Index
Voltage: 11.1VDC
Rated Battery capacity: 4.4 Ah (one battery)
Battery run time: >270min (two batteries), >130 min (one battery)
Operating conditions: 25ºC for new fully-charged battery. Equipment configuration: 5-lead
ECG, Impedance Repiration, Masimo SpO
2
, Temperature, IBP1, IBP2, and NIBP in 15-min
Interval mode.
Battery charge time:
3.5 hours for 90% charge and 5.5 hours for full charge (equipment powered off)
5.5 hours for 90% charge and 8.5 hours for full charge (equipment powered on)
0

Table of Contents

Other manuals for Mindray Passport V

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Passport V and is the answer not in the manual?

Mindray Passport V Specifications

General IconGeneral
BrandMindray
ModelPassport V
CategoryHealth Care Products
LanguageEnglish