Prompt Messages Messages
14 - 10 0070- 0-0704-02 Passport V Operating Instructions
14.3.4 Temp Prompt Message
SpO
2
: Interference Patient Movement.
External interference.
Noise detected on
the pulse signal
prevents pulse
discrimination.
Decrease patient
motion,check sensor.
Contact Technical
Support.
Masimo/Nellcor
SpO
2
: No Sensor No sensor Apply sensor Masimo/DPM/
Nellcor
SpO
2
: Weak Pulse Low pulse amplitude Check the condition of
poor perfusion for
patient.
Contact Technical
Support.
Nellcor
SpO
2
: Low Signal The SpO
2
signal is
too low or too weak.
Check sensor
placement, move as
necessary.
Switch limb / Notify
physician.
Nellcor
SpO
2
: Weak Signal The SpO
2
signal is
too low or too weak.
Check patient
Reposition sensor
Change sensor
Masimo
SpO
2
: Check Sensor Possible Faulty
sensor
Replace Sensor Nellcor
SpO
2
: Motion Patient movement Check patient Nellcor
SpO
2
: Pulse Search Monitor searching
for pulse.
Sensor just applied
Poor perfusion at
sensor site.
Check patient.
Reposition sensor.
Masimo/DPM/
Nellcor
ALARM MESSAGE REASON SOLUTION
Temp: Communication Stop As the Temp module
communication stops, the data
packets sent by the Temp module
can not be received.
Contact Technical Support.
Temp: SelfTest Error Module send out error during
selftest because of some
unspecified reasons.
Contact Technical Support.
Temp: Calibration Error A calibration failed. Restart the monitor.
Contact Technical Support.
Temp: Initialization Error During the Temperature module
power-on,as the Temperature
module communication
stops,system fails to
communicate with module.
Contact Technical Support.
0