Passport V Operating Instructions 0070- 0-0704-02 4 - 29
ECG Monitoring Arrhythmia Analysis (Optional)
The following alarm calls will be made when Arrhythmia analysis is set to “Non-lethals Off”:
• Asystole, Ventricular Tachycardia, and Ventricular Fibrillation.
When Arrhythmia analysis is set to “All Off,” no arrhythmia alarm calls will be made.
NOTE: The “All Off” selection in the “Arrhythmia Menu” is only
available if “Enable Arrhythmia All Off Selection” is set to
“Yes” in the Installation Menu.
4.8.1 Arrhythmia Analysis Setup
ARR TYPE OPTION DESCRIPTION
Asystole Asystole Delay No QRS complex is detected within the set "Asystole Delay",
which is judged as one Asystole event.
V-Tach V-Tach Rate
V-Tach PVC
HR is greater than or equal to the set "V-Tach Rate" and the
number of continuous PVCs is greater than or equal to the set
"V-Tach PVC" threshold, which is judged as one V-Tach
event.
Brady Brady Low When HR is 10% less than the HR low alarm limit , it is
judged as one Bradycardia event.
0