EasyManuals Logo
Home>Mindray>Health Care Products>Passport V

Mindray Passport V Operating Instructions

Mindray Passport V
288 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #171 background imageLoading...
Page #171 background image
Alarms Alarms
13 - 4 0070- 0-0704-02 Passport V Operating Instructions
HIGH LOW
PARAMETERS ADULT PED NEONATE ADULT PED NEONATE
HR (bpm)
Except France
150, Off,
60-250
180, Off,
100-300
200, Off,
100-350
50, Off,
30-120
80, Off,
30-150
100, Off,
30-200
HR (bpm)
France
120, Dés,
60-250
150, Dés,
100-300
175, Dés,
100-350
30, Dés,
30-120
50, Dés,
30-150
70, Dés,
30-200
ST Single
Lead (mm)
Off,
+0.5 to
+10.0
Off,
+0.5 to
+10.0
Off,
-0.5 to
-10.0
Off,
-0.5 to
-10.0
ST Dual
Lead (mm)
Off,
+0.5 to
+10.0
Off,
+0.5 to
+10.0
Off,
-0.5 to
-10.0
Off,
-0.5 to
-10.0
SpO
2
(%) Off,
80-100
Off,
80-100
Off,
80-100
85,
76-99
90,
76-99
92,
76-99
Desat (%) 80,
75-98
80,
75-98
80,
75-98
Resp (RPM) Off,
10-100
Off,
15-150
Off,
30-200
Off,
5-30
Off,
5-40
Off,
5-50
NIBP Sys (mmHg) Off,
70-240
Off,
40-180
Off,
40-180
Off,
50-150
Off,
15-130
Off,
15-130
NIBP Dia (mmHg) Off,
40-130
Off,
50-100
Off,
50-100
Off,
30-120
Off,
10-50
Off,
10-50
NIBP Mean
(mmHg)
Off,
60-200
Off,
50-180
Off,
40-160
Off,
40-140
Off
,
10-100
Off,
10-70
Temp (ºF) Off,
95-110
Off,
95-110
Off,
95-110
Off,
80-100
Off,
80-100
Off,
80-100
Temp (ºC) Off,
35-43
Off,
35-43
Off,
35-43
Off,
26-38
Off,
26-38
Off,
26-38
IBP1 Sys (mmHg) Off,
5-300
Off,
5-240
Off,
5-180
Off,
0-150
Off,
0-130
Off,
0-130
IBP1 Dia (mmHg) Off,
0-140
Off,
0-100
Off,
0-70
Off,
0-120
Off,
0-100
Off,
0-50
IBP1 Mean
(mmHg)
Off,
5-150
Off,
5-100
Off,
5-100
Off,
2-100
Off,
2-50
Off,
2-50
IBP2 Sys (mmHg) Off,
5-300
Off,
5-240
Off,
5-180
Off,
0-150
Off,
0-130
Off,
0-130
IBP2 Dia (mmHg) Off,
0-140
Off,
0-100
Off,
0-70
Off,
0-120
Off,
0-100
Off,
0-50
IBP2 Mean
(mmHg)
Off,
5-150
Off,
5-100
Off,
5-100
Off,
2-100
Off,
2-50
Off,
2-50
ET CO
2
(mmHg) 60, Off,
20-80
60, Off,
20-80
60, Off,
20-80
Off,
5-50
Off,
5-50
Off,
5-50
ET CO
2
(%) 8.0, Off,
2.0-10.0
8.0, Off,
2.0-10.0
8.0, Off,
2.0-10.0
Off,
1.0-6.0
Off,
1.0-6.0
Off,
1.0-6.0
ET CO
2
(kPa) 8.0, Off,
2.0-10.0
8.0, Off,
2.0-10.0
8.0, Off,
2.0-10.0
Off,
1.0-6.0
Off,
1.0-6.0
Off,
1.0-6.0
Insp CO
2
(mmHg) 10
Off, 5-30
10
Off, 5-30
10
Off, 5-30
——
Default values in bold.
0

Table of Contents

Other manuals for Mindray Passport V

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Passport V and is the answer not in the manual?

Mindray Passport V Specifications

General IconGeneral
BrandMindray
ModelPassport V
CategoryHealth Care Products
LanguageEnglish