Passport V Operating Instructions 0070- 0-0704-02 16 - 3
Printing Print Setup Menu
Print on Alarm
If a waveform is initiated by an alarm violation, then the local printer will initiate a 16-second
or 32-second waveform strip. The print length (16 or 32-seconds) can be set in the Print
Setup menu. To activate this feature, from the front panel, select Alarms: Limits
Print On
Alarm Active. Then select each parameter for which the feature should be enabled.
During an alarm printout, the data for the parameter which violated the alarm will be marked
with an asterisk. The alarm trigger for the printout is indicated on the strip with the name of
the violated alarm.
Interval Printing
To print waveform strips at regular intervals, go to the Print Setup Menu, highlight Print
Interval and select a time interval to print. The selections are Off, 1 min, 5 min, 10
min, 15 min, 20 min, 30 min, 1 hr and 2 hr. The source of the printout is indicated on
the strip as AUTO REALTIME REPORT.
Trend Printing
To print Quick Trend, List Trend, OxyCRG, or Graphic Trend data to the local printer,
press the Print Trend key while the trend is displayed. To abort printing, press the Cont
ECG, Strip, or Print Trend key.
Quick Trend and List Trend print digital data information while Graphic Trend print graphic
representations of the data in the respective menus. Quick Trend entries are only recorded
upon NIBP readings.
FIGURE 16-2 Graphic Trend Format
FIGURE 16-3 List Trend Format
0

Table of Contents

Other manuals for Mindray Passport V

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Passport V and is the answer not in the manual?

Mindray Passport V Specifications

General IconGeneral
BrandMindray
ModelPassport V
CategoryHealth Care Products
LanguageEnglish