Passport V Operating Instructions 0070- 0-0704-02 19 - 27
Appendix Performance Characteristics
Rise times (t10–90%)
3
@120 ml/min
• CO
2
250 ms (fall time 200 ms)
• N
2
O 250 ms
• O
2
600 ms
• HAL, ISO, SEV, DES 300 ms
• ENF 350 ms
Delay time
4
(t0—10%) < 4 s
Identification Dual agent
Primary agent ID threshold
5
0.15% (0.4% during ISO accuracy mode)
Secondary agent ID threshold
5
0.3% (0.5% during ISO accuracy mode) or 5% REL
(10% REL for Isoflurane) of primary agent if primary
agent >10%
Agent ID time Three breaths - Typically less than 10 s
Display Update Rate Breath-by-Breath
Compensation The Gas Module 3 is equipped with automatic barometric
pressure compensation.
Main Fuse 2x T0.8A 250V
1 Includes interference from other gases.
2 The step rise time specification at 200 ml/min sample flow includes DRYLINE• Water Trap, Adult/Pediatric
and DRYLINE• Sampling Line, Adult/Pediatric 2.5 m.
3 The step rise time specification at 120 ml/min sample flow includes DRYLINE• Water Trap, Neonate and
DRYLINE• Sampling Line, Neonate 2.5 m.
4 The delay time specification is valid both for 120 ml/min sample flow (using DRYLINE• Water Trap, Neo-
nate and DRYLINE• Sampling Line, Neonate 2.5 m) and for 200 ml/min sample flow (using DRYLINE•
Water Trap, Adult/Pediatric and DRYLINE• Sampling Line, Adult/Pediatric 2.5 m).
5 For HAL, add 0.1% ABS to threshold values.
0