Passport V Operating Instructions 0070- 0-0704-02 4 - 35
ECG Monitoring ECG Troubleshooting
Base Line Wander Patient moving excessively. Secure lead wires and cable to
patient.
Patient's respiration. Reposition electrodes.
Electrodes dry or loose. Re-prep skin and apply fresh,
moist electrodes.
Static build up around patient. Check with local biomedical
personnel.
ECG Filter set to “ST” or
“Diagnostic” mode.
Set ECG Filter to “Monitor”
mode.