Mode Sense Command
122 T10000: Interface Reference Manual August 2016
SOCF: Stop On Consecutive Filemarks 00b = Stop read ahead when buffer is full
RBO: Recover Buffer Order 0 = Not supported
REW: Report Early Warning 0 = Report early warning only on Write and Write
Filemarks commands
Gap Size 0 = Gap size not selectable
EOD Defined: End Of Data 000b = Default EOD only
EEG: EOD Enabled Generation 1 = EOD generated per EOD field
SEW: Synchronize at Early Warning (LEOT) 0 = Buffered write data and filemarks not flushed
to the tape when LEOT is detected
1 = Buffered write data and filemarks written to
the tape when LEOT is detected
SWP: Soft Write Protect 0 = Not supported
Buffer Size at Early Warning 0 = Buffer size not selectable
Select Algorithm:
Select Data Compression Algorithm
Default is operator configurable
00h = No data compression
01h = LZ1 compression of write records
ASOCWP: Associated Write Protect 0 = Not supported
PERSWP: Persistent Write Protect 0 = Not supported
PRMWP: Permanent Write Protect 0 = Not supported
Parameters Value