6.3 Physical Layer Specifications
6.3.1 Communication Options
Host communication options such as baud rate can be found on the Utility / System / Host
Communication Setting screen.
Item Specifications Remarks
Communication Speed
[bps]
1200 , 2400 , 4800 , *9600
19200, 38400, 57600, 115200
Baud rate
Handshake
OFF
HW
*SW
No handshake
RTS, CTS mode
Xon, Xoff mode
Parity Check
*OFF
ON
Parity check disabled
Parity check enabled
Line Mode *N_8_1
E_8_1
O_8_1
N_8_2
E_7_1
O_7_1
E_7_2
O_7_2
None Parity, 8 bit, 1 stop bit
Even Parity, 8 bit, 1 stop bit
Odd Parity, 8 bit, 1 stop bit
None Parity, 8 bit, 2 stop bits
Even Parity, 7 bit, 1 stop bit
Odd Parity, 7 bit, 1 stop bit
Even Parity, 7 bit, 2 stop bits
Odd Parity, 7 bit, 2 stop bits
Communication Port 1
Electrical Signal In accordance with EIA-232-D-1986
Cable Length Maximum 15m
Table 3: Physical Level Specifications